Safety
•High accuracy: ±2%
•Fast start function against start delay
•Anti-bolus function
•DERS (Drug error reduction system)
•IP Level: IP34
•11 hours battery back-up time
Friendliness
•4.3" colour touch screen
•Intuitive operation system
•Automatic calibration
•Inter-Lockable
Specifications and Functions
Dimension242 * 126* 111
WeightApprox.1.7kg
Display4.3 inch colour touch screen
Syringe sizes2ml,5ml, 10ml,20ml,30ml,50/60ml; Automatic recognition
Flow rate accuracy±2%
Flow rate0.1-1500 ml/h (depending on syringe size, with increment 0.01 ml/h)
VTBI0-9999.99 ml
Dose rate unitsMore than 15 types
Concentration calculateAutomatically
Bolus settingManual bolus Programmable bolus
KVO rate0.1-5.0 ml/h
Infusion modes8 infusion modes
(Rate mode, Time mode, Body-weight mode, TIVA mode, Drug library mode, Ramp up/down mode,Loading-dose mode, Sequence mode)
HandleIncluded
Drug libraryMore than 2000
PurgeYes
DPSYes
TitrationYes
Micro modeYes
Standby modeYes
DERSYes
Screen lockYes
Occlusion levels12 levels
Anti-bolusAutomatically
RecordsMore than 5000 entries
AlarmsVTBI near end, VTBI infused, Pressure high,
Battery nearly empty, Battery empty, No power supply, Reminder alarm, KVO finished, Check syringe, Syringe empty, Standby time expired, System error
Safety
Power supplyAC: 100V-240V, 50/60Hz DC: 12V
Battery lifeStandard: 5.5 hours; Optional: 11 hours (@5ml/h)
Charging time< 5 hours
ClassificationClass I, CF
IP LevelIP34
Interface
lrDAOptional
Data interfaceUSB
WirelessWiFi (optional )
DC inputYes
RS232/
Nurse callSupported