Infusion pumpAMK321

Main features Reliable and Safe:

Advanced safety design able to stop infusion and alarm even if the CPU damage,the medical staff can ues it safely.

Applicable syringe 5ml,10ml,20ml,30ml,50/60ml

The maximum flow rate range 5ml syringe: 0.1ml/h~200ml/h

10ml syringe: 1ml/h~300ml/h

20ml syringe: 1ml/h~400ml/h

30ml syringe: 1ml/h~600ml/h

50/60ml syringe: 1ml/h~1200ml/h

Precision Better than ±2%;

(but 5ml syringe and <1ml/h: Better than ±8%)

Parameter range 1, Flow-rate: 0.1~1200.0 ml/h, 0.1ml/h step;

2, Dose: 0.1~999.9 mg, 0.1mg step;

3, mg/kg/h: 0.001~999.999 mg/kg/h, 0.001 mg/kg/h step;

4, Body-weight: 1~300 kg, 1kg step;

5, KVO rate: 0.1~5.0 ml/h, 0.1 ml/h step;

6, Bolus rate: 0.1~1200.0 ml/h, 0.1 ml/h step;

7, Bolus volume: 0.1~99.9 ml/h, 0.1ml step

Alarm 1, Occlusion,0~175KPa, 8 grades of sensitivity;

2, Pressure sensor is abnormal;

3, Push rod fall off;

4, The drug running out;

5, The injection will finish;

6, The motor rotation fault;

7, Overtime pause;

8, Low battery;

9, Battery failure;

10, Battery temperature is too high;

11, Mains missing tips;

12, MCU fault, and stop injection.

Data function 1, Memory of parameter: Records the latest injection parameter for next operation;

2, Historical records: records latest 1000 injection events.

(what/when/why,flow-rate, etc.)

Safety classification Class I; Internal power supply, CF type, IPX4 equipment

Power supply AC 100~240V, 50/60Hz, Built-in rechargeable Li-battery

Power consumption:20VA

Working time of battery: 10h@5ml/h

Weight and size 37cm(w)*24cm(h)*21cm(t); about 2.1kg

Optional accessories Wireless communicating module.

Chinese Certificate yes